SPECIFICATION

Field of View: Fixed 70H x 25W x 40D[mm] cuboid
Real-Time Data Rate: ≤ 150Hz, platform dependent. Full system from acquisition to output
Strain Outputs: Analog Output: ±10V, short-protected, selectable units and range; includes 3m (10ft) shielded output cable
Resolution: <0.5 μm quasistatic, and cyclic; Typical RMS resolution at typical settings. Resolution is a function of the marking data rate and filter settings
Extensometer Accuracy Class: ISO 9513 Class 0,5 and ASTM E83 Class B-1 capable
Minimum Specimen Width/Diameter: 2mm wide flat or 2.5mm diameter round, excluding marking accessories
Gage Length: 10-50mm
Maximum Tracking Speed: >2500mm/min
Cyclic Testing: Speed and specimen dependent
Strain Control: Suitable for monotonic and cyclic strain control applications
Out-of-Plane Sensitivity: Out-of-plane boundary = 280 to 320mm from front of module
Ambient Operating Light Conditions: Suitable for use in a day light situation, artificial lab or room lighting situations
Suitable for use with Temperature
Chambers:
Measuring strain in a temperature chamber at ambient, high and low temperature (min. window width 170mm)
Signals Integration Options: Control, information, and notification through client interface and or test suite software. Manufacturer dependent
Power Supply: 100-240 VAC, 50-60 Hz, 1.4A 120W max, IEC 320 C14 receptacle.
Power Consumption: 8W (avg)
Recycling Capability: Up to 60% of this product can be recycled; metals and cabling
System Environment: 10° to 40°C (50 to 100°F), for use and storage; 20-80% relative humidity noncondensing environments
Dimensions: 252H x 201W x 73D mm
Mechanical Integration Options: Physical mounting schemes available for;
T slot single and twin column vertical testing machine frames.
Multi-column vertical testing machine frames.
Horizontal testing machine frames.
Typical ROI: 6 months
Component Defect Warranty: 12 months
Manufactured: In the UK